Description
Specifications
Product model | IPRM-G-NACLO |
Measurement item | Refractive Index (nD), Brix,temperature. |
Measuring range | 0.00 – 15 .00% Brix |
Resolution | 0.01% Brix |
Measurement accuracy | ± 0.2% Brix |
Brix range | Temperature corrected 5 – 50°C |
Process temperature5 | -10°C up to +100°C |
Prism | Sapphire |
Wetted parts | HastelloyC-276 |
Power supply | 24 V DC |
Interfaces | 1 output 4 – 20 mA,1 serial output RS485 |
International Protection class | Display surface:IP65 Water resistant |
Process pressure | Max. 1 MPa (145 psi, 10 bar) |
Dimensions and weight | length 138 x diam.109mm; approx. 1870g |
Cleaning components Options | IPRM-US, IPRM-Wiper |
Fitting Options | IDF/ISO clamp union (1-3 inches), Flange, Compression Fitting G1/4 |
Dimensions
Measurement Principles
The refraction of light is a phenomenon wherein the direction that light is traveling changes as it passes through different mediums.
Through this phenomenon, we can observe as light’s direction changes and the “bent” appearance of substances even in the seemingly familiar, ordinary places in our day to day lives.
For example, imagine the lens of a pair of glasses and a pool.
When looking through the lens, an object appears larger or smaller than it is in reality.
When standing inside a pool, as you peer downward from the water’s surface, your legs will look shorter than they actually are.
As light moves from medium A (in this case, air) to medium B (lens and the pool’s water), it refracts.
Furthermore, it is known that light’s refraction changes depending on the concentration of a liquid.
The refractive index, which is a value that expresses the degree of refraction, is proportional to the density (mixture ratio) of a component.
Therefore, if there is 10g of sugar contained in 100ml of an aqueous solution and 20g of sugar contained in 100ml of an aqueous solution, the latter will have a higher refractive index value.
This difference in refractive index due to concentration variance has also been confirmed in solutions other than sucrose.
The properties of both the phenomenon known as the refraction of light and differing refractive index depending on concentration, are used as the principles of refractometers.
In addition to measuring the refractive index of a liquid, refractometers also display the concentration, converted from the refractive index.
Product introduction
IPRM-G-NACLO is a precision instrument for production process monitoring with high precision and high temperature resistance, and an inline refractometer to ensure the normal operation of product quality and production process, helping enterprises improve production efficiency, enhance automation, reduce labor costs, reduce production costs, and ensure the consistency and stability of product quality in the production process.
Product features
IPRM-G-NACLO sodium hypochlorite concentration process refractometer adopts the principle of critical angle refractive index detection, which can be directly installed on the wall of pipes or tank vessels with different pipe diameters, and the detection is not affected by impurities such as pressure, flow rate, color, turbidity, bubbles, solid particles and so on; With high-precision automatic temperature compensation function, shielding anti-interference, automatic cleaning.
Product advantages
● It is not affected by the color, turbidity and viscosity of the material to be measured
● Not affected by bubbles, solid impurities and crystals in the material
● It is not affected by liquid pressure changes, flow changes, turbulent flow phenomena
● No mechanical action measurement, ensure stability and long life
● The detection prism is an optical grade sapphire prism, which is resistant to scratches, scratches and abrasion
● Detection of prism full-plane structure; No reagent consumables are required to run, and low power consumption
● With automatic temperature compensation function, the measurement accuracy is not affected by temperature
● Integrated measurement and display; Support panel and Bluetooth communication settings, view parameters
● 4-20mA/RS485; Built-in alarm, cleaning and switching signal output
● With 15 years of industry application experience, Chuyi Measurement and Control provides customers with the most complete product application solutions
● Chuyi Measurement and Control’s laboratory can provide data modeling and testing services for customers’ specific solutions
Application highlights
IPRM-G-NACLO Sodium hypochlorite process refractometer is usually used in the disinfection and sterilization process monitoring of industrial water treatment.